7
T
ROUBLESHOOTING
7.14
ThinPrep
®
3000 Processor Operator’s Manual
Insufficient Vial Fluid
Reprocessing Required
The vial does not contain enough sample fluid to properly process the sample. The sample was not
processed and a slide was not made.
Possible Cause
• Previous testing with the sample vial may have reduced the volume of sample fluid in the vial
below the minimum required.
• A displaced or missing cap liner may have caused fluid to leak out of the sample vial.
• A pneumatic system failure has occurred.
WARNING:
Toxic Mixture
Flammable Liquid and Vapor
Corrective Action
1. The batch report indicates the position of the vial under the Reprocessing Required listing as
Insufficient Vial Fluid. Check the fluid level in the vial.
2. Add additional PreservCyt Solution to the vial if fluid level is below the frosted area indicated on
the vial. Do not fill beyond the frosted area.
3. Reprocess the sample vial.
4. If error recurs with the sample vial, perform the pneumatic system test. Refer to “PNEUMATIC
Preventative Action
Prior to sample processing, ensure that the sample fluid level is within the frosted area of the sample
vial.
ID#
00001
168
Frosted area
Содержание ThinPrep 3000
Страница 1: ......
Страница 4: ...The ThinPrep Processor The ThinPrep Processor ...
Страница 5: ...MAN 03939 001 Rev 004 page 1 of 13 Instructions for Use ...
Страница 18: ...Table of Contents Table of Contents ...
Страница 23: ...1 Introduction 1 Introduction ...
Страница 41: ...1 INTRODUCTION 1 18 ThinPrep 3000 Processor Operator s Manual This page intentionally left blank ...
Страница 42: ...2 ThinPrep 3000 Installation 2 ThinPrep 3000 Installation ...
Страница 54: ...2 THINPREP 3000 INSTALLATION 2 12 ThinPrep 3000 Processor Operator s Manual This page intentionally left blank ...
Страница 55: ...3 PreservCyt and CellFyx Solutions 3 PreservCyt and CellFyx Solutions ...
Страница 71: ...4 Sample Collection and Preparation 4 Sample Collection and Preparation ...
Страница 80: ...5 Instrument Operation 5 Instrument Operation ...
Страница 105: ...6 Maintenance 6 Maintenance ...
Страница 147: ...6 INSTRUMENT MAINTENANCE 6 42 ThinPrep 3000 Processor Operator s Manual This page intentionally left blank ...
Страница 148: ...7 Troubleshooting 7 Troubleshooting ...
Страница 204: ...7 TROUBLESHOOTING 7 56 ThinPrep 3000 Processor Operator s Manual This page intentionally left blank ...
Страница 205: ...8 Staining and Coverslipping 8 Staining and Coverslipping ...
Страница 211: ...8 STAINING AND COVERSLIPPING 8 6 ThinPrep 3000 Processor Operator s Manual This page intentionally left blank ...
Страница 212: ...9 The ThinPrep Pap Test Training Program 9 The ThinPrep Pap Test Training Program ...
Страница 215: ...10 User Interface Screens 10 User Interface Screens ...
Страница 226: ...Index Index ...
Страница 232: ...INDEX Index 6 ThinPrep 3000 Processor Operator s Manual This page intentionally left blank ...
Страница 233: ...Service Information Service Information ...
Страница 236: ...Ordering Information Ordering Information ...
Страница 242: ...ORDERING INFORMATION Ordering 6 ThinPrep 3000 Processor Operator s Manual This page intentionally left blank ...
Страница 243: ...Safety Data Sheets Safety Data Sheets ...
Страница 245: ...SAFETY DATA SHEETS SDS 2 ThinPrep 3000 Processor Operator s Manual This page intentionally left blank ...
Страница 246: ...Appendix Appendix ...
Страница 255: ......
Страница 256: ......