Pressing the button, the maximum and minimum values obtained during the test are displayed. The maximum and minimum values
are reset when the test function is changed or the meter is switched off.
6) Relative
The instrument displays relative measurement results in all functions except the Hz frequency.
7] Hz% Frequency
During the Hz% measurement, pressing this button will switch the meter between the HZ function and the work cycle function.
8) Function selection knob
With this switch you can select the needed function and range.
9) Socket „A”
For red (positive) lead (200mA ~ 10A).
10) COM socket
for the black (negative) lead.
11) INPUT socket
Plug-in connector of the red lead for all measurements except for current measurements (≥200mA).
5. SPECIFICATION
5-1. DC voltage (Auto ranging)
Range
Resolution
Accuracy
400mV
0.1mV
±(0.8% r 5 digits)
4V
1mV
±(0.8% r 3 digits)
40V
10mV
400V
100mV
600V
1V
±(1.0% r 5 digits)
Input impedance: 10 MΩ
Overload protection: 600 V DC or 600 AC RMS
(Range 200 mV: 250 V DC / AC RMS)
Max. Input voltage: 600 V
5-2. AC voltage (Auto ranging)
Range
Resolution
Accuracy
400 mV
1 mV
±(1.2% r 5 digits)
4 V
1 mV
±(1.0% r 5 digits)
40 V
10 mV
400 V
100 mV
600 V
1 V
±(1.2% r 8 digits)
Input impedance: 10 MΩ
Frequency range: 40 Hz ~ 400 Hz
Overload protection: 600 V DC or 600 AC RMS
Answer: Average, calibrated in rms sinusoidal wave
Max. Input voltage: 600 V AC RMS
5-3. TEMPERATURE
Range
Resolution
Accuracy
-40 ~1370°C
1°C
-40°C~150°C:±(2.5% + 4) about
150°C~1370°C:±(2.5% + 4)
-40 ~2000°F
1°F
-40°F~302°F:±(2.5% + 4) about
302°F~2000°F:±(2.5% + 4)
NOTICE: Different temperature sensors are configured in different temperature test ranges, and standard temperature sensors
are designed for standard configuration