HOBO MX pH and Temperature Logger (MX2501) Manual 

U.S. and International Sales: 1-508-759-9500 9 

www.onsetcomp.com 

 

There are two mounting options: 

 

Attach a nylon cord or other strong cable to the 
mounting hole to deploy the logger. 

 

 

Alternatively, secure the logger to a pipe or similar 
surface using cable ties wrapped around the two large 
grooves in the logger housing.  

 

 

The logger should be appropriately secured and 
protected depending on water conditions and desired 
measurement location. 

Maintaining the Logger 

 

The pH sensor must be kept in storage solution when 
not being calibrated or deployed in water

. Additional 

storage solution (MX2500-STORE-SOLN), clear storage 
caps (MX2500-STORAGE-CAP), and a maintenance kit 
containing pH 4.01, 7.00, and 10.00 calibration solution 
and storage solution (MX2500-MAINT-KIT) are available. 
See www.onsetcomp.com. 

 

Periodically inspect the logger for biofouling. To clean the 
logger housing, rinse the logger in warm water. Use mild 
dishwashing soap if necessary on the logger housing only, 
taking care not to touch the glass bulb in the pH 
electrode. Do not use harsh chemicals, solvents, or 
abrasives. Replacement sensor end caps (MX2500-END-
CAP) and copper guards (MX2500-GUARD) are available if 
they cannot be cleaned effectively. See 
www.onsetcomp.com for replacement parts. 

 

Periodically uninstall the pH electrode from the logger to 
inspect it for deposits. Clean and recondition it as follows 
depending on the type of deposits found. To remove a 
build-up of scale or other dissolve salts, briefly soak the 
pH sensor in 5% hydrochloric acid. To remove oil 
deposits, soak the pH sensor in a solution with mild 
dishwashing soap. 

Never touch the glass bulb inside the 

clear plastic end of the probe. 

Always calibrate the 

logger after cleaning the pH electrode with anything 
other than distilled or deionized water. 

 

The hydration layer surrounding the glass pH sensor bulb 
in the electrode naturally breaks down over time. 
Cleaning and reconditioning the electrode as described in 

in the previous bullet can help to extend its life. It is 
recommended that you replace the pH electrode after 6 
months of continuous use or when the slope and offset 
readings reported during calibration are consistently not 
within an acceptable range, even after cleaning. The ideal 
range for slope is 85% to 100% and for offset is ±30 mv. 
See 

Installing the pH Electrode

 for steps on replacing the 

electrode (MX2500-ELECTRODE).  

 

Periodically add silicone-based grease to the outer O-
rings on the sensor end cap shown below. Using your 
finger, spread a small dot of grease along both outer O-
rings. 

 

Installing the pH Electrode 

Important:

 New pH electrodes are uncalibrated. Once a pH 

electrode is installed in the logger for the first time, it must be 
calibrated before configuring and deploying the logger (see 

Calibrating the Logger

). 

To install the pH electrode in the logger: 

1.

 

Unscrew the closure cap over the sensor end cap of the 
logger. Pull the sensor end cap straight out from the logger 
without twisting it. 

 

2.

 

Gently and slowly

 remove the pH electrode from the 

sensor end cap. (

Note:

 Removing the electrode too quickly 

could result in unseating the internal O-rings.) 

 

Use this hole for 
mounting the logger 

Outer O-rings on 
sensor end cap 

Use cable ties around these 
grooves for mounting the logger 

First, unscrew this closure cap 

Then, pull this 
sensor end cap 
straight out 

Gently and slowly

 

remove the pH electrode 
from the end cap 

Содержание MX2501

Страница 1: ...ater Deionized water or water below 100 S cm may not be detected The water conductivity circuit may not reliably detect water that has frozen around the electrodes i e below 0 C 32 F Radio Power 1 mW...

Страница 2: ...n Union EU See last page Logger Components and Operation Anti Biofouling Copper Guard Use this guard to protect the pH electrode when deploying the logger in water prone to biofouling see Deploying an...

Страница 3: ...your mobile device or computer The range for successful wireless communication is approximately 30 5 m 100 ft in air with full line of sight Change the orientation of your device to ensure the antenn...

Страница 4: ...2 If you removed the clear storage cap screw the closure cap onto the logger 3 In the HOBOconnect app tap Devices 4 Remove the logger from water If the logger was previously configured with Bluetooth...

Страница 5: ...rt below of buffers and samples Temperature C F pH 4 01 Buffer pH 7 00 Buffer pH 10 00 Buffer 0 32 4 01 7 12 10 31 5 41 4 00 7 09 10 24 10 50 4 00 7 06 10 17 15 59 4 00 7 04 10 11 20 68 4 00 7 02 10 0...

Страница 6: ...owing Cumulative The alarm trips once the sensor reading is out of the acceptable range for the selected duration any time during logging For example if the high alarm is set to 85 F and the duration...

Страница 7: ...can be logged for each sensor The maximum or highest sampled value The minimum or lowest sampled value An average of all sampled values The standard deviation from the average for all sampled values...

Страница 8: ...bration End pH calibration ended while the logger was logging Calibration Cancel pH calibration was canceled while the logger was logging Calibration Values The slope and offset values that were set d...

Страница 9: ...solution with mild dishwashing soap Never touch the glass bulb inside the clear plastic end of the probe Always calibrate the logger after cleaning the pH electrode with anything other than distilled...

Страница 10: ...logger immediately screw the closure cap back on the logger and follow the instructions in Calibrating the Logger If you will not be calibrating or deploying the logger immediately a Pour storage sol...

Страница 11: ...8 Reinstall the battery holder and attached mounting end cap Use the alignment tab on the battery holder shown in step 4 and the arrow on the logger housing as a guide as you fit the small tab in the...

Страница 12: ...ny interference received including interference that may cause undesired operation FCC Caution Any changes or modifications not expressly approved by the party responsible for compliance could void th...

Отзывы: