background image

HOBO MX pH and Temperature Logger (MX2501) Manual 

U.S. and International Sales: 1-508-759-9500 4 

www.onsetcomp.com 

sure you have the time to complete the 
entire update process, which requires that 
the logger remains connected to the device 
during the upgrade. 

 

Calibrating the Logger

 

If the logger is new or if you have installed a new pH electrode, 
you must calibrate the logger with the app before configuring 
and deploying it. The app prompts you to calibrate the logger if 
the pH electrode has never been calibrated, the last calibration 
is more than seven days old, or the logger firmware was 
updated. 

Important:

 

The pH sensor must be kept in storage solution 

when not being calibrated or deployed in water

. The hydration 

layer surrounding the sensor starts to break down after 10 
minutes if the pH sensor is not submerged in storage solution 
or calibration solution, or deployed in water. 

You will need: 

 

Deionized or distilled water and a squirt bottle to rinse 
the pH sensor. 

 

The logger with the pH electrode installed and the copper 
guard removed. 

 

Calibration solution (also called buffer solution) poured 
into beakers. All three buffers are available in a kit 
(MX2500-CAL-KIT) or individually (MX2500-CAL-4, 
MX2500-CAL-7, or MX2500-CAL-10). 

Important:

 The logger should be calibrated based on the 

conditions at the location where it will be deployed. 
Specifically, if the water temperature is expected to be 
different than 25°C (77°F), it is important for the calibration 
buffers to be close to the expected temperature of the analyte 
(see the chart at the end of this section for the effect of 
temperature on buffers). Similarly, if the logger will be 
deployed at an altitude other than sea level, perform the 
calibration at or near the deployment altitude. 

To calibrate the logger: 

1.

 

If the sensor is in the storage solution, unscrew the clear 
storage cap and remove it from the logger.  

 

2.

 

If you removed the clear storage cap, screw the closure cap 
onto the logger. 

 

3.

 

In the HOBOconnect app, tap Devices. 

4.

 

Remove the logger from water. If the logger was previously 
configured with Bluetooth Always Off, press the switch on 
the logger to wake it up.  

5.

 

Tap the logger in the app to connect to it. 

6.

 

Tap Calibrate if you are not automatically prompted to 
calibrate the logger. 

7.

 

Use a squirt bottle to rinse the pH sensor with deionized or 
distilled water. 

8.

 

Place the sensor end of the logger in the pH 7 solution and 
tap Start. Make sure the sensor end cap, temperature 
sensor, and closure cap are submerged in the solution as 
shown in the illustration. Once the pH reading is stable, tap 
Confirm. 

 

9.

 

Rinse the pH sensor with deionized or distilled water. 

10.

 

If you are using the pH 4 buffer, place the sensor end in the 
pH 4 solution and tap Start. If you are not using the pH 4 
buffer, tap Skip and go to step 12. 

11.

 

Once the pH reading is stable, tap Confirm. Rinse the pH 
sensor with deionized or distilled water. 

12.

 

If you are using the 10 buffer, place the sensor end in the 
pH 10 solution and then tap Start. Once the pH reading is 
stable, tap Confirm. If you are not using the pH 10 buffer, 
tap Skip. 

13.

 

Verify that the slope and offset are acceptable and tap 
Save. If the slope and offset are not acceptable, tap the 
back arrow and repeat the calibration steps in this section 
as needed. 

14.

 

Rinse the pH sensor with deionized or distilled water again. 

If you are not deploying the logger immediately, place the pH 
electrode in storage solution

. Unscrew the sensor end cap 

installed on the sensor end of the logger. If the clear storage 
cap is empty, pour in storage solution to the marked fill line. 
Carefully screw the clear cap with storage solution onto the 
logger. Only use storage solution in the clear cap; do not use 
water. Do not use the copper guard when the pH electrode is in 
storage solution. 

Unscrew the clear 
cap filled with 
storage solution 
from the logger 

Screw the closure 
cap onto the logger 
before calibrating 

Closure cap 

Sensor end cap 

Temperature sensor 

Содержание MX2501

Страница 1: ...ater Deionized water or water below 100 S cm may not be detected The water conductivity circuit may not reliably detect water that has frozen around the electrodes i e below 0 C 32 F Radio Power 1 mW...

Страница 2: ...n Union EU See last page Logger Components and Operation Anti Biofouling Copper Guard Use this guard to protect the pH electrode when deploying the logger in water prone to biofouling see Deploying an...

Страница 3: ...your mobile device or computer The range for successful wireless communication is approximately 30 5 m 100 ft in air with full line of sight Change the orientation of your device to ensure the antenn...

Страница 4: ...2 If you removed the clear storage cap screw the closure cap onto the logger 3 In the HOBOconnect app tap Devices 4 Remove the logger from water If the logger was previously configured with Bluetooth...

Страница 5: ...rt below of buffers and samples Temperature C F pH 4 01 Buffer pH 7 00 Buffer pH 10 00 Buffer 0 32 4 01 7 12 10 31 5 41 4 00 7 09 10 24 10 50 4 00 7 06 10 17 15 59 4 00 7 04 10 11 20 68 4 00 7 02 10 0...

Страница 6: ...owing Cumulative The alarm trips once the sensor reading is out of the acceptable range for the selected duration any time during logging For example if the high alarm is set to 85 F and the duration...

Страница 7: ...can be logged for each sensor The maximum or highest sampled value The minimum or lowest sampled value An average of all sampled values The standard deviation from the average for all sampled values...

Страница 8: ...bration End pH calibration ended while the logger was logging Calibration Cancel pH calibration was canceled while the logger was logging Calibration Values The slope and offset values that were set d...

Страница 9: ...solution with mild dishwashing soap Never touch the glass bulb inside the clear plastic end of the probe Always calibrate the logger after cleaning the pH electrode with anything other than distilled...

Страница 10: ...logger immediately screw the closure cap back on the logger and follow the instructions in Calibrating the Logger If you will not be calibrating or deploying the logger immediately a Pour storage sol...

Страница 11: ...8 Reinstall the battery holder and attached mounting end cap Use the alignment tab on the battery holder shown in step 4 and the arrow on the logger housing as a guide as you fit the small tab in the...

Страница 12: ...ny interference received including interference that may cause undesired operation FCC Caution Any changes or modifications not expressly approved by the party responsible for compliance could void th...

Отзывы: