background image

HOBO MX pH and Temperature Logger (MX2501) Manual 

U.S. and International Sales: 1-508-759-9500 6 

www.onsetcomp.com 

9.

 

Tap Pause Options, then select Pause On Button Push to 
specify that you can pause the logger by pressing its switch 
for 3 seconds. 

10.

 

Tap Logging Mode. Select either Fixed Logging Mode or 
Burst Logging. With fixed logging, the logger records data 
for all sensors and/or selected statistics at the logging 
interval selected (see 

Statistics Logging

 for details on 

choosing statistics options). In burst mode, logging occurs at 
a different interval when a specified condition is met. See 

Burst Logging

 for more information. 

11.

 

Enable or disable Show LED. If Show LED is disabled, the 
alarm and status LEDs on the logger are not illuminated 
while logging (the alarm LED does not blink if an alarm 
trips). You can temporarily turn on LEDs when Show LED is 
disabled by pressing the button on the logger for 1 second. 

12.

 

Select the power saving mode, which determines when the 
logger advertises or regularly sends out a Bluetooth signal 
for the mobile device or computer to find via the app. 

 

Bluetooth Always Off.

 The logger advertises only during 

logging when you push the switch on the logger. This 
wakes up the logger when you need to connect to it. This 
option uses the least battery power. 

 

Bluetooth Always On.

 The logger always advertises. You 

never need to push a switch to wake up the logger. This 
option uses the most battery power. 

13.

 

Set up an optional alarm to trip when the pH and/or 
temperature sensor reading rises above or falls below a 
specified value. See 

Setting up Alarms

 for details on 

enabling sensor alarms. 

14.

 

Tap Start to save the configuration settings. 

Logging begins based on the settings you selected. See 

Deploying and Mounting the Logger

 for details on mounting 

and see 

Reading Out the Logger 

for details on downloading.

 

Setting up Alarms 

You can set up alarms for the logger so that if a sensor reading 
rises above or falls below a specified value, the red logger LED 
blinks and an alarm icon appears in the app. Alarms alert you to 
problems so you can take corrective action. 

To set an alarm: 

1.

 

Tap Devices. If the logger was configured with Bluetooth 
Always Off enabled, push the switch on the logger to wake 
it up. 

2.

 

Tap the logger tile to connect to the logger and tap 
Configure & Start. 

3.

 

Tap a Sensor/Channel. 

4.

 

Tap Enable Logging, if necessary. 

5.

 

Tap Alarms to open that area of the screen. 

6.

 

Select Low if you want an alarm to trip when the sensor 
reading falls below the low alarm value. Enter a value to set 
the low alarm value. 

7.

 

Select High if you want an alarm to trip when the sensor 
reading rises above the high alarm value. Enter a value to 
set the high alarm value. 

8.

 

For the Duration, select how much time should elapse 
before the alarm trips and select one of the following: 

 

Cumulative. The alarm trips once the sensor reading is 
out of the acceptable range for the selected duration any 

time during logging. For example, if the high alarm is set 
to 85°F and the duration is set to 30 minutes, the alarm 
trips once the sensor readings have been above 85°F for 
a total of 30 minutes since the logger was configured. 

 

Consecutive. The alarm trips once the sensor reading is 
out of the acceptable range continuously for the selected 
duration. For example, the high alarm is set to 85°F and 
the duration is set to 30 minutes, the alarm trips only if 
all sensor readings are 85°F or above for a continuous 30-
minute period. 

When an alarm trips, the red logger LED blinks every 4 seconds 
(unless Show LED is disabled), an alarm icon appears in the app, 
and an Alarm Tripped event is logged. The alarm state clears 
when the readings return to normal if you selected Sensor in 
Limits in step 9. Otherwise, the alarm state remains in place 
until the logger is reconfigured. 

Notes: 

 

Alarm limits are checked at every logging interval unless 
burst logging is configured. For example, if the logging 
interval is set to 5 minutes, the logger checks the sensor 
readings against your configured high and low alarm 
setting every 5 minutes. 

 

The actual values for the high and low alarm limits are set 
to the closest value supported by the logger. For 
example, the closest value to 85°F that the logger can 
record is 84.990°F. In addition, alarms can trip or clear 
when the sensor reading is within the specified 
resolution specifications. 

 

When you read out the logger, alarm events can be 
displayed on the plot or in the data file. See 

Logger 

Events

Burst Logging 

Burst logging is a logging mode that allows you to set up more 
frequent logging when a specified condition is met. For 
example, a logger is recording data at a 5-minute logging 
interval and burst logging is configured to log every 30 seconds 
when the temperature rises above 85°F (the high limit) or falls 
below 32°F (the low limit). This means the logger records data 
every 5 minutes as long as the temperature remains between 
85°F and 32°F. Once the temperature rises above 85°F, the 
logger switches to the faster logging rate and records data 
every 30 seconds until the temperature falls back to 85°F. At 
that time, logging then resumes every 5 minutes at the fixed 
logging interval. Similarly, if the temperature falls below 32°F, 
then the logger switches to burst logging mode again and 
record data every 30 seconds. Once the temperature rises back 
to 32°F, the logger then returns to fixed mode, logging every 5 
minutes. 

Note:

 Sensor alarms, statistics, and the Stop Logging 

option Never Stop (Overwrites Old Data) are not available in 
burst logging mode. 

To set up burst logging: 

1.

 

Tap Devices. Remove the logger from water. If the logger is 
configured with Bluetooth Always Off enabled, press the 
button on the logger to wake it up.  

2.

 

Tap the logger tile to connect to the logger and tap 
Configure & Start. 

3.

 

Tap Logging Mode and then tap Burst Logging. 

Содержание MX2501

Страница 1: ...ater Deionized water or water below 100 S cm may not be detected The water conductivity circuit may not reliably detect water that has frozen around the electrodes i e below 0 C 32 F Radio Power 1 mW...

Страница 2: ...n Union EU See last page Logger Components and Operation Anti Biofouling Copper Guard Use this guard to protect the pH electrode when deploying the logger in water prone to biofouling see Deploying an...

Страница 3: ...your mobile device or computer The range for successful wireless communication is approximately 30 5 m 100 ft in air with full line of sight Change the orientation of your device to ensure the antenn...

Страница 4: ...2 If you removed the clear storage cap screw the closure cap onto the logger 3 In the HOBOconnect app tap Devices 4 Remove the logger from water If the logger was previously configured with Bluetooth...

Страница 5: ...rt below of buffers and samples Temperature C F pH 4 01 Buffer pH 7 00 Buffer pH 10 00 Buffer 0 32 4 01 7 12 10 31 5 41 4 00 7 09 10 24 10 50 4 00 7 06 10 17 15 59 4 00 7 04 10 11 20 68 4 00 7 02 10 0...

Страница 6: ...owing Cumulative The alarm trips once the sensor reading is out of the acceptable range for the selected duration any time during logging For example if the high alarm is set to 85 F and the duration...

Страница 7: ...can be logged for each sensor The maximum or highest sampled value The minimum or lowest sampled value An average of all sampled values The standard deviation from the average for all sampled values...

Страница 8: ...bration End pH calibration ended while the logger was logging Calibration Cancel pH calibration was canceled while the logger was logging Calibration Values The slope and offset values that were set d...

Страница 9: ...solution with mild dishwashing soap Never touch the glass bulb inside the clear plastic end of the probe Always calibrate the logger after cleaning the pH electrode with anything other than distilled...

Страница 10: ...logger immediately screw the closure cap back on the logger and follow the instructions in Calibrating the Logger If you will not be calibrating or deploying the logger immediately a Pour storage sol...

Страница 11: ...8 Reinstall the battery holder and attached mounting end cap Use the alignment tab on the battery holder shown in step 4 and the arrow on the logger housing as a guide as you fit the small tab in the...

Страница 12: ...ny interference received including interference that may cause undesired operation FCC Caution Any changes or modifications not expressly approved by the party responsible for compliance could void th...

Отзывы: