background image

HOBO MX pH and Temperature Logger (MX2501) Manual 

U.S. and International Sales: 1-508-759-9500 7 

www.onsetcomp.com 

4.

 

Set the burst logging interval, which must be faster than the 
logging interval. Keep in mind that the faster the burst 
logging rate, the greater the impact on battery life and the 
shorter the logging duration. Because measurements are 
being taken at the burst logging interval throughout the 
deployment, the battery usage is similar to what it would be 
if you had selected this rate for the normal logging interval. 

5.

 

Select Low and/or High and type a value to set the low 
and/or high values. Select Low and/or High and either type 
or drag the slider to set the low and/or high values. 

Notes: 

 

The high and low burst limits are checked at the burst 
logging interval rate whether the logger is in normal or 
burst condition. For example, if the logging interval is set 
to 1 hour and the burst logging interval is set to 10 
minutes, the logger always checks for burst limits every 
10 minutes. 

 

If high and/or low limits have been configured for more 
than one sensor, then burst logging begins when any high 
or low condition goes out of range. Burst logging does 
not end until all conditions on all sensors are back within 
normal range. 

 

The actual values for the burst logging limits are set to 
the closest value supported by the logger. In addition, 
burst logging can begin or end when the sensor reading is 
within the specified resolution. This means the value that 
triggers burst logging may differ slightly than the value 
entered.  

 

Once the high or low condition clears, the logging 
interval time is calculated using the last recorded data 
point in burst logging mode, not the last data point 
recorded at the fixed logging rate. For example, the 
logger has a 10-minute logging interval and logged a data 
point at 9:05. Then, the high limit is surpassed and burst 
logging begins at 9:06. Burst logging then continues until 
9:12 when the sensor reading falls back below the high 
limit. Now back in fixed mode, the next logging interval is 
10 minutes from the last burst logging point, or 9:22 in 
this case. If burst logging had not occurred, the next data 
point would have been at 9:15. 

 

A New Interval event is created each time the logger 
enters or exits burst logging mode. See 

Logger Events

 for 

details on plotting and viewing the event. In addition, if 
the logger is stopped with a button push while in burst 
logging mode, a New Interval event is automatically 
logged and the burst condition is cleared, even if the 
actual high or low condition has not cleared. 

Statistics Logging 

During fixed interval logging, the logger records data for 
enabled sensors and/or selected statistics at the logging 
interval selected. Statistics are calculated at a sampling rate you 
specify with the results for the sampling period recorded at 
each logging interval. The following statistics can be logged for 
each sensor: 

 

The maximum, or highest, sampled value 

 

The minimum, or lowest, sampled value 

 

An average of all sampled values 

 

The standard deviation from the average for all sampled 
values 

For example, a logger is configured with the logging interval set 
to 5 minutes and Fixed Logging Mode. In addition, all four 
statistics are enabled with a statistics sampling interval of 30 
seconds. Once logging begins, the logger measures and records 
the actual temperature and pH sensor values every 5 minutes. 
In addition, the logger takes a temperature and pH sample 
every 30 seconds and temporarily store them in memory. The 
logger then calculates the maximum, minimum, average, and 
standard deviation using the samples gathered over the 
previous 5-minute period and log the resulting values. When 
downloading data from the logger, this results in 10 data series 
(not including the mV series): two sensor series (with 
temperature and pH data logged every 5 minutes) plus eight 
maximum, minimum, average, and standard deviation series 
(four for temperature and four for pH with values calculated 
and logged every 5 minutes based on the 30-second sampling). 

To log statistics: 

1.

 

Tap Devices. Remove the logger from water. If the logger 
was configured with Bluetooth Always Off enabled, press 
the button on the logger to wake it up.  

2.

 

Tap the logger tile in the app to connect to the logger and 
tap Configure & Start. 

3.

 

Tap Logging Mode and then select Fixed Logging Mode. 

4.

 

Select the statistics you want the logger to record at each 
logging interval: Maximum, Minimum, Average, and 
Standard Deviation (average is automatically enabled when 
selecting Standard Deviation). Statistics are logged for all 
enabled sensors. In addition, the more statistics you record, 
the shorter the logger duration and the more memory is 
required. 

5.

 

Tap Statistics Sampling Interval and select the rate to use 
for calculating statistics. The rate must be less than, and a 
factor of, the logging interval. For example, if the logging 
interval is 1 minute and you select 5 seconds for the 
sampling rate, the logger takes 12 sample readings between 
each logging interval (one sample every 5 seconds for a 
minute) and use the 12 samples to record the resulting 
statistics at each 1-minute logging interval. Note that the 
faster the sampling rate, the greater the effect on battery 
life. Because measurements are being taken at the statistics 
sampling interval throughout the deployment, the battery 
usage is similar to what it would be if you had selected this 
rate for the normal logging interval. 

Setting a Password 

You can create an encrypted password for the logger that is 
required if another device attempts to connect to it. This is 
recommended to ensure that a deployed logger is not 
mistakenly stopped or purposely altered by others. This 
password uses a proprietary encryption algorithm that changes 
with every connection. 

To set a password: 

1.

 

Tap Devices. Press the switch 

on the logger

 to wake it up, if 

necessary. Tap the logger tile in the app to connect to it. 

2.

 

Tap Lock Logger. 

3.

 

Type a password and then tap Set. 

Содержание MX2501

Страница 1: ...ater Deionized water or water below 100 S cm may not be detected The water conductivity circuit may not reliably detect water that has frozen around the electrodes i e below 0 C 32 F Radio Power 1 mW...

Страница 2: ...n Union EU See last page Logger Components and Operation Anti Biofouling Copper Guard Use this guard to protect the pH electrode when deploying the logger in water prone to biofouling see Deploying an...

Страница 3: ...your mobile device or computer The range for successful wireless communication is approximately 30 5 m 100 ft in air with full line of sight Change the orientation of your device to ensure the antenn...

Страница 4: ...2 If you removed the clear storage cap screw the closure cap onto the logger 3 In the HOBOconnect app tap Devices 4 Remove the logger from water If the logger was previously configured with Bluetooth...

Страница 5: ...rt below of buffers and samples Temperature C F pH 4 01 Buffer pH 7 00 Buffer pH 10 00 Buffer 0 32 4 01 7 12 10 31 5 41 4 00 7 09 10 24 10 50 4 00 7 06 10 17 15 59 4 00 7 04 10 11 20 68 4 00 7 02 10 0...

Страница 6: ...owing Cumulative The alarm trips once the sensor reading is out of the acceptable range for the selected duration any time during logging For example if the high alarm is set to 85 F and the duration...

Страница 7: ...can be logged for each sensor The maximum or highest sampled value The minimum or lowest sampled value An average of all sampled values The standard deviation from the average for all sampled values...

Страница 8: ...bration End pH calibration ended while the logger was logging Calibration Cancel pH calibration was canceled while the logger was logging Calibration Values The slope and offset values that were set d...

Страница 9: ...solution with mild dishwashing soap Never touch the glass bulb inside the clear plastic end of the probe Always calibrate the logger after cleaning the pH electrode with anything other than distilled...

Страница 10: ...logger immediately screw the closure cap back on the logger and follow the instructions in Calibrating the Logger If you will not be calibrating or deploying the logger immediately a Pour storage sol...

Страница 11: ...8 Reinstall the battery holder and attached mounting end cap Use the alignment tab on the battery holder shown in step 4 and the arrow on the logger housing as a guide as you fit the small tab in the...

Страница 12: ...ny interference received including interference that may cause undesired operation FCC Caution Any changes or modifications not expressly approved by the party responsible for compliance could void th...

Отзывы: