5.3
Manual Cleaning and disinfection
Prepare following items before manual cleaning and disinfection:
a)
Detergent: Cidezyme® (Johnson & Johnson, #2258) or another cleaning
agent with approved material compatibility for this medical device.
b)
Disinfectant: Cidex® OPA (Johnson & Johnson, # 20391) or another
disinfectant with approved material compatibility for this medical
device
c)
Cleaning brushes if applicable
d)
Two tanks, one for cleaning and one for disinfection - optional:
1 additional tank for rinsing with deionized/tap water (sufficient
size for immersion of the submergible part of the probe at full
length)
e)
Soft, fluff free cloth or single use towel
f)
50 ml syringe
g)
Personal protective equipment (gloves, water repellent protective
skirt, face protection mask or protective glasses, see also
instructions of the manufacturer for the detergent and the
disinfectant)