-11-
Q1E-EP1412
CAUTION
3.2.1
Fitting the needle guide bracket to the probe
1)
Insert the recess of needle guide bracket to the needle guide bracket
mount part. (See Fig. 6)
2)
Fit the recess of the bracket to the thumbscrew 1 and tighten the
thumbscrew 1. (See Fig. 7) At this time, be sure to attach the needle
guide bracket so that the needle guide attachment area is positioned
in the same direction as that of the needle guide bracket mount part
side.
Never attach the needle guide attachment area at the reverse side of
the needle guide bracket mount part. Otherwise, the biopsy guide line
becomes inconsistent with the inserting position of the biopsy needle.
Needle guide
bracket
Needle guide
bracket mount part
Needle guide
attachment area
Fig. 6 Position for mounting the needle guide bracket
Thumbscrew 1
Fig. 7 Fitting the needle guide bracket
CAUTION