-8-
Q1E-EP1412
4)
Use a sterile probe cover to protect the probe. The probe cover
should be allergy free material to avoid allergic reaction. Between
the probe and the probe cover, acoustic coupling gel is required
as a couplant.
5)
Place the probe and the Magnetic Sensor Attachment on the examination
site and adjust the probe’s position for a clear view of the desire
image.
6)
After using the probe, perform the reprocessing procedure in
accordance with the procedure stated in “4. Reprocessing Procedure”
every time immediately after completing the ultrasound examination.
7)
Store the probe, the Magnetic Sensor Attachment and the Spacer for
EZU-RV2S in the environment indicated in “5. Maintenance and Safety
Inspection”.
3.2
Needle Guide Bracket EZU-PA7C2
1)
Confirm that the needle guide bracket is cleaned and disinfected,
and the probe is cleaned and disinfected/ sterilized.
2)
Fit the needle guide bracket to the probe following “3.2.1 Fitting
the needle guide bracket to the probe”. Never apply excessive force
to attach the needle guide bracket to the wrong position. It may
cause a hazard due to unstable biopsy.
3)
Fit the CIV-Flex™ cover to the probe following “3.2.2 Placing the
probe and bracket into a transducer cover”.
4)
Attach the needle guide to the bracket following “3.2.3 Attaching
the needle guide to the bracket”.
5)
Fit the needle insert corresponding to the gauge number of the needle
to be used to the needle guide following “3.2.3 Attaching the needle
guide to the bracket”.
6)
Confirm that the needle guide angle corresponds to the angle
indicated on the monitor. The needle guide angle is engraved on
bracket following “3.2.4 Setting the needle guide angle”.
7)
In the case of the CIVCO Ultra Pro II, when the needle is released
quickly from probe, press the tab on the needle insert toward the
bracket. Move bracket and needle guide away from the needle. (See
Fig.5)
8)
After the use of the needle guide bracket, it should be cleaned and
disinfected, then store it in an adequate place.
Содержание EUP-B712
Страница 48: ... 42 Q1E EP1412 Unit mm Fig 26 Dimensions ...
Страница 49: ... 43 Q1E EP1412 Unit mm Fig 27 External view with the Magnetic Sensor Attachment ...
Страница 50: ... 44 Q1E EP1412 Unit mm Fig 28 Dimensions with the Needle Guide Bracket ...
Страница 51: ... 45 E Q1E EP1412 Unit mm Fig 29 External view with Ultra Pro II Needle Guide cover kit ...