Point of use (Pre-cleaning)
Pre-cleaning should be done immediately after each
use. The procedure is as follows:
1)
Remove the EZU-PA3U from the probe.
2)
Disassemble the EZU-PA3U (Fig.14) and immerse all parts in
sufficient amount of high quality tap water. Scrub them using soft
cloth to remove all visible soil and dried protein from their
surface.
6.2
Containment and transportation
Refer to 5.2.
6.3
Manual Cleaning and disinfection
Prepare following items before manual
cleaning and disinfection:
a)
Detergent: Cidezyme (Johnson & Johnson,
#2258) or another cleaning agent with
approved material compatibility for this
medical device
b)
Disinfectant: Cidex OPA (Johnson & Johnson,
# 20391) or another disinfectant with
approved material compatibility for this
medical device
c)
Applicable brush
Fig.14 Disassembly of EZU-PA3U
Point of use
(Pre-cleaning)
Manual Cleaning
Rinsing after manual
cleaning
Manual Disinfection
Rinsing after manual
disinfection
Manual cleaning and
disinfection
Containment and
transportation