- 31 -
Q1E-EP1451
1)
The parameters of the cleaning and disinfection of the device are
as follows:
Program step Water (40 l) Dosage (ml/l) Temp. (°C) Time (min)
Pre-Rinse Cold
water
-
-
5
Cleaning Deionized water
5 (0.5%)
50
5
Rinse Deionized
water
-
-
1
Disinfection Deionized water
10 (1%)
55
5
Rinse Deionized
water
-
-
1
Rinse Deionized
water
-
55
1
Drying -
- 55
15
2)
Place the disassembled parts of EZU-PA3U in the baskets of the washer
disinfector. Smaller parts like the screws must be placed in a
strainer basket with lid.
3)
Close the door of the washer disinfector and start the chemo-thermal
program.
4)
Open the door after the process is done.
5)
Take the parts of EZU-PA3U out of the washer disinfector and check
that they are dry. If not, dry them as described in the chapter drying.
6.5
Drying
Refer to 5.4.
6.6
Inspection
Refer to 5.5.
6.7
Packaging
Refer to 5.6.
Drying
Packaging