- 27 -
Q1E-EP1451
6. Cleaning, Disinfection and Sterilization of EZU-PA3U
EZU-PA3U must be reprocessed after each use. Refer to the
reprocessing instruction in this chapter.
WARNINGS
EZU-PA3U is delivered unsterile. Prior to the first
use, reprocess it.
Transportation
before using
EZU-PA3U should be packed in a sterile pouch or
container to transport from Central Sterile Supply
Department (CSSD) to an operating room. Be careful
not to damage the sterile pouch or container during
transportation.
Levels of reprocessing requirements:
Refer to the chapter 5.
The flowchart of the reprocessing process of EZU-PA3U is as follows.
Point of use (Pre-cleaning)
Manual Cleaning
Rinsing after manual
cleaning
Manual Disinfection
Drying
Automated Cleaning
Manual cleaning and
disinfection
Washer Disinfector (WD)
Automated Disinfection
Rinsing after manual
disinfection
Packing
Sterilization
Containment and transportation