5-2. Disinfection
Either chemical disinfection or gas disinfection is performed as necessary.
5-2-1. Chemical disinfection
Applicable chemicals
General name
Trade name
Manufacturer
Glutaral
CIDEX™
Solution 2.4%
ADVANCED STERILIZATION PRODUCTS
®
A Johnson & Johnson company
Division of Ethicon, Inc.
Ortho-phthalaldehyde
CIDEX OPA™
Solution 0.55%
Glutaral
STERIHYDE™
Practical liquid 2W/V%
Maruishi Pharmaceutical Co., Ltd.
Benzalkonium chloride
DETERGICIDE™
Practical liquid 0.2W/V%
Yufu Itonaga Co., Ltd.
Benzethonium chloride
HYAMINE™
Practical liquid 0.1W/V%
DAIICHI SANKYO Co., Ltd.
Disinfection procedure
Washing
Be sure to wash and fully dry the scanner before disinfection.
See section 5-1.
Soaking in
chemical solution
Soak the insertion portion and handle in a chemical solution at the
shortest time where effective disinfection can be achieved.
Water rinsing
Sufficiently rinse the scanner using sterilized water to remove
chemicals.
Remarks
Soaking the insertion portion and handle in CIDEX OPA™ solution 0.55% may result in discoloration of the
silicone, but this does not affect performance or safety.
Warning
After chemical disinfection, thoroughly wash the scanner with sterilized water.
Residual chemicals can cause an adverse reaction on the bodies of the operator or patient.
Содержание ASU-1003
Страница 2: ...MN1 5008 Rev 7 ii ...
Страница 6: ...MN1 5008 Rev 7 vi ...
Страница 14: ... 8 MN1 5008 Rev 7 Label A 2 Storage case Label B ...
Страница 22: ... 16 MN1 5008 Rev 7 ...
Страница 34: ... 28 MN1 5008 Rev 7 ...
Страница 36: ... 30 MN1 5008 Rev 7 ...
Страница 38: ... 32 MN1 5008 Rev 7 ...
Страница 42: ... 36 MN1 5008 Rev 7 ...