4-2. Mounting of rubber boot
Apply a suitable amount of acoustic medium to the ultrasonic irradiation area of the scanner and then cover
it with the rubber boot.
Remove the air bubbles or wrinkles form on the ultrasonic irradiation area of the scanner.
Apply a suitable amount
of acoustic medium.
Rubber boot
Warning
Use by covering the rubber boot over the insertion portion.
If the rubber boot is not used, residual pathogens on the scanner could infect the patient. Also,
the puncture guide tube could become loose during puncturing, resulting in puncturing of an
unintended body part.
Use Aloka-approved rubber boots only.
Use of an item lacking biocompatibility can cause an adverse reaction by the body of the patient.
Check that the rubber boot is sterilized.
Use of an infected item could spread infection to the patient.
Do not reuse the rubber boot.
Use of an infected item could spread infection to the patient.
Do not apply unsterilized acoustic medium to the outer surface of the rubber boot.
Use of an acoustic medium that is contaminated by a pathogen can cause an infection on the
patient.
Do not use on patients who may have an allergic reaction to latex products.
Use of the rubber boot for these types of patients could result in anaphylactic shock. Ask the patient
about allergy history beforehand.
Caution
Check the rubber boot for abnormalities before use.
Store the rubber boots in a cool, dry location not exposed to direct sunlight and do not use rubber
boots that have exceeded their expiration date (for items where the expiration date is not displayed,
2 years from the displayed sterilization date) or severe discoloration, cracks, or other visible defects
finds.
Check that the acoustic medium has no air bubbles inside the rubber boot that is covering the
scanner.
Air bubbles inside the rubber boot can result in misdiagnosis caused by overlooking or
misinterpreting lesions due to poor image quality or improper rendering.
Содержание ASU-1003
Страница 2: ...MN1 5008 Rev 7 ii ...
Страница 6: ...MN1 5008 Rev 7 vi ...
Страница 14: ... 8 MN1 5008 Rev 7 Label A 2 Storage case Label B ...
Страница 22: ... 16 MN1 5008 Rev 7 ...
Страница 34: ... 28 MN1 5008 Rev 7 ...
Страница 36: ... 30 MN1 5008 Rev 7 ...
Страница 38: ... 32 MN1 5008 Rev 7 ...
Страница 42: ... 36 MN1 5008 Rev 7 ...