52
Read Error Information
Syntax
Query
:SYSTem:ERRor?
Response
<Error No.(NR1)>
<Error No.> =
32768
16384
8192
4096
2048
1024
512
256
bit 15
bit 14
bit 13
bit 12
bit 11
bit 10
bit 9
bit 8
RAM error
RSV
RSV
RSV
Mode error
AD error
Measurement
error
RSV
128
64
32
16
8
4
2
1
bit 7
bit 6
bit 5
bit 4
bit 3
bit 2
bit 1
bit 0
RSV
Configuration
error
Backup
error
Storage
memory
error
RSV
MAC address
error
Adjustment
value
error
ROM
error
Description
Queries the error number.
Error contents
Details →
Corrective Action
ROM error
An operation error occurred in the internal ROM.
→ The repair is needed.
Adjustment value error
The adjustment value showed an abnormal
value.
→ The repair is needed.
MAC address error
The MAC address showed an abnormal value.
This may cause the communication failure to
occur.
→ The repair is needed.
Storage memory error
The operation of the memory for storage of
settings was abnormal.
→ The repair is needed.
Backup error
The instrument was started with the default
settings as the stored settings showed an
abnormal value.
→ When the error is cleared by restarting the
instrument, there is no problem.
Configuration error
The configuration at startup failed.
→ The repair is needed.
AD error
The AD value showed an abnormal value.
→ The repair is needed.
Mode error
The measurement might be executed in an
inappropriate state.
→Please inform the details of the processing
to the nearest office.
Measurement error
The sampling failed during measurement.
→ The repair is needed.
RAM error
An operation error occurred in the internal RAM.
→ The repair is needed.
Example
:SYST:ERR?
48
B
The storage error and backup error occurred.