42
Query Measured Values of Tristimulus Values XYZ (
# is R, G, B, or RGB
)
Syntax
Query
:FETCh:XYZ:#?
Response
<X measured value(NR3)>,<Y measured value(NR3)>,<Z measured
value(NR3)>,<Measurement status(NR1)>
See:
“Measurement Value Formats” (p.37), “Measurement Status” (p.36)
Description
Reads the last measured values of the tristimulus values XYZ.
Example
:MODE NORM
:READ?
*TRG
3.7109E-01,3.4633E-01,4.03,0
:FETC:XYZ:R?
3.03,1.03,1.72926E-01,0
X, Y, and Z of the red tristimulus values XYZ are 3011.97, 1211.05, and 0.172926,
respectively.
:FETC:XYZ:G?
9.02,2.03,6.01,0
X, Y, and Z of the green tristimulus values XYZ are 904.522, 2957.30, and 62.2899,
respectively.
:FETC:XYZ:B?
6.02,8.01,3.03,0
X, Y, and Z of the blue tristimulus values XYZ are 636.569, 80.9570, and 3404.54,
respectively.
:FETC:XYZ:RGB?
4.03,4.03, 3.03,0
X, Y, and Z of the tristimulus values of the RGB mixed light are 4553.06, 4249.23, and
3467.00, respectively.
Note
• # is R, G, B, or RGB.
• Be sure to check that the measurement status is an appropriate value.
• If the measurement mode is not the normal measurement mode, an execution error
occurs.