41
Query Measured Value of Radiometric Quantity (
# is R, G, or B
)
Syntax
Query
:FETCh:RADiometry:#?
Response
<Measured value of radiometric quantity(NR3)>,<Measurement status(NR1)>
See:
“Measurement Value Formats” (p.37), “Measurement Status” (p.36), “Unit of
Photometric Quantity” (p.9)
Description
Reads the last measured value of the radiometric quantity.
Example
(For TM6102)
:MODE NORM
:READ?
*TRG
3.7109E-01,3.4633E-01,4.03,0
:FETC:RAD:R?
7.00,0
The red radiometric quantity is 7.92924 [W/m
2
].
:FETC:RAD:G?
4.00,0
The green radiometric quantity is 4.53508 [W/m
2
].
:FETC:RAD:B?
2.00,0
The blue radiometric quantity is 2.82641 [W/m
2
].
:FETC:RAD:RGB?
1.01,0
The radiometric quantity of the RGB mixed light is 15.2907 [W/m
2
].
Note
• # is R, G, B, or RGB.
• Be sure to check that the measurement status is an appropriate value.
• If the measurement mode is not the normal measurement mode, an execution error
occurs.