25
Query Set Number of Average Times of Dark Measurement
Syntax
Command
:DARK:AVERaging
<Average Times (NRf)>
Query
:DARK:AVERaging?
Response
<Average times(NR1)>
<Average times>
=
1 to 100
Description
Command
Sets the number of average times of the dark measurement.
When the command is accepted, the measured value is cleared and the
instrument enters the non-measurement state.
Query
Queries the currently set number of average times of the dark measurement.
Example
:DARK:AVER 2
:DARK:AVER?
2
Query Judgment Function for Dark Measurement Result
Syntax
Command
:DARK:JUDGment
<1/0/ON/OFF >
Query
:DARK:JUDGment?
Response
<1/0(NR1)>
Description
Command
Sets the judgment function for the dark measurement result.
• ON is set.
The response of :READ?
or
:FETCh:DARK?
is 1(PASS) or 0(FAIL) in
accordance with the dark measurement environment.
• OFF is set.
The response of :READ?
or
:FETCh:DARK?
is always 1 (the dark
measurement result is the PASS judgment).
When the command is accepted, the measured value is cleared and the
instrument enters the non-measurement state.
Query
Queries the currently set judgment function for the dark measurement result.
Example
:DARK:JUDG ON
:DARK:JUDG?
1
Note
• When this setting is set to OFF, the normal measurement may not be performed
correctly if the dark measurement is performed in bright environments.