47
Query Detection Levels
Syntax
Query
:FETCh:LEVel?
Response
<Red detection level(NR2)>,<Green detection level(NR2)>,<Blue detection
level(NR2)>
<Detection level of each color> = Unit: [%]
See:
“Measurement Value Formats” (p.37)
Description
Queries the last measured detection levels.
The detection level shows what percentage of the F.S. of the measured range has been
measured.
For example, when the detection level is 20%, the measurement is performed with 20% of
the F.S. Therefore, it is better to perform the measurement by increasing the sensitivity of
the measurement range.
Example
:MODE NORM
:READ?
*TRG
3.7209E-01,3.4709E-01,1.03,0
:FETC:LEV?
40.60,40.70,56.83
Note
• When the manual range is used, use this command to select a range in which the
detection level is maximum.
• If the measurement mode is not the normal measurement mode, an execution error
occurs.
Query Modulation Frequency Measurement Result
Syntax
Query
:FETCh:PULSe?
Response
<Modulation frequency(NR2)>,<Measurement status(NR1)>
<Modulation frequency> = Unit: [Hz]
See:
“Measurement Value Formats” (p.37), “Measurement Status” (p.36)
Description
Reads the last measured value of the modulation frequency.
Example
:MODE PULS
:READ?
*TRG
60.0097,0
:FETC:PULS?
60.0097,0
Note
• If the measurement mode is not the modulation frequency measurement mode, an
execution error occurs.
Query Dark Measurement Result
Syntax
Query
:FETCh:DARK?
Response
<Judgment result(NR1)>
<Judgment result> = 1/0 1: PASS judgment (Dark measurement is succeeded.)
0: FAIL judgment (Dark measurement is failed.)
Description
Reads the dark measurement result measured last.
Example
:MODE DARK
:READ?
*TRG
1
:FETC:DARK?
1
Note
• If the measurement mode is not the dark measurement mode, an execution error occurs.