49
:FETC:RAD:B?
7.00,0
:TARG:RES:B?
8.00,0,1,5.00,1.01
Measured value of blue radiometric quantity: 7.48744 [W/m
2
]
Target value of blue radiometric quantity: 8.36588 [W/m
2
]
Lower threshold of tolerance: 5.05005 [W/m
2
] Upper threshold of
tolerance: 12.7150 [W/m
2
]
Judgment: PASS judgment (The measured value of the radiometric quantity is between
the upper and lower thresholds of the tolerance.)
Note
• # is R, G, B, or RGB.
• Be sure to check that the measurement status is an appropriate value.
• If the measurement mode is not the modulation frequency measurement mode, an
execution error occurs.
• The information acquired by this command should be handled as a reference value. The
information that is used for the final judgment is acquired by
:TARGet:RESult:RGB?
.
• In rare cases, even when the measured values of the radiometric quantities of all colors
are between the upper and lower thresholds of the tolerance of the radiometric quantity,
the response of
:TARGet:RESult:RGB?
may be the FAIL judgment. In this case, perform the adjustment
so that the radiometric quantity of each color gets close to the adjustment target of the
radiometric quantity.
• The floating-point operation is performed inside the instrument. The judgment is
performed including fractions after the response digits.
The value after the display digits is rounded off. Therefore, even when the measurement
result is the same as the upper or lower threshold of the adjustment target, the judgment
result may vary.