21
Connecting the Connection Cable
TERMINAL 3
1
Turn off the device.
2
Connect the connection cable to
TERMINAL 3
.
Front
Match the cable and the connection
terminal signal.
SOURCE-H (red)
SENSE-H (red)
SENSE-L (black)
SOURCE-L (black)
1. Check the direction of
the grooves of the BNC
connector. The grooves
should fit into the connector
guide of the device.
2. Align the grooves of the
BNC connector with the
connector guide of the
device and insert the BNC
connector.
3. Turn the BNC connector to
the right to lock it.
Device side
Connector guide
BNC connector grooves of
the connection cord
3
Connect the other end of the connection cable to the instrument.
Connection cable terminal
Instrument terminal
BT4560
IM3590
SOURCE-H (red)
SOURCE-H (red)
Hcur
SENSE-H (red)
SENSE-H (red)
Hpot
SENSE-L (black)
SENSE-L (black)
Lpot
SOURCE-L (black)
SOURCE-L (black)
Lcur
2
Preparation for Measurements