137
SW9002 Multiplexer Module (4-terminal pair)
Measurement influence
(1)
Conditions
Warm-up time
None
Temperature and
humidity
23°C±5°C (73°F±9°F), 80% RH or less
Measurement
conditions
Measurement cable length 0 m (Connect the measuring object to the SW9002's
connector.)
Use the specified connection cable between the switch mainframe and the target
measuring instrument.
Based on the conditions separately defined for each target measuring instrument.
Based on the guaranteed accuracy conditions for use in conjunction with an instrument.
(2)
Combined measurement accuracy
The combined measurement accuracy for an instrument and the switch (main frame and module)
can be calculated using the following formula:
Combined measurement accuracy = Instrument ac combination measurement influence
(3)
Combination measurement influence
BT4560 (connected using L2004)
Range
Measurement influence
Conditions, remarks
Frequency range
0.1 Hz to 100 Hz
Frequency range
110 Hz to 1050 Hz
3 m
Ω
R
±0.05% f.s.
±0.1% f.s.
–
3 m
Ω
X
±0.1% f.s.
±1.0% f.s.
–
10 m
Ω
R
±0.015% f.s.
±0.03% f.s.
–
10 m
Ω
X
±0.03% f.s.
±0.3% f.s.
–
100 m
Ω
R
±0.01% f.s.
±0.01% f.s.
–
100 m
Ω
X
±0.015% f.s.
±0.03% f.s.
–
All V ranges
±5 μV*
1
After the ambient operating temperature is
stabilized.
Within 1 minute after the contact is closed.
IM3590*
2
(connected using L2004)
Range
Measurement influence
Conditions, remarks
100 m
Ω
to 10
Ω
(IM3590 measurement accuracy) × 1
DC, 0.001 Hz to 10.000 kHz
100
Ω
to 10 k
Ω
(IM3590 measurement accuracy) × 3
DC, 0.001 Hz to 10.000 kHz
Upper limit of impedance 10 k
Ω
DM7275 and DM7276 (connected using L4930)
Range
Measurement influence
Conditions, remarks
All V ranges
±7 μV*
1
After the ambient operating temperature is
stabilized.
Within 1 minute after the contact is closed.
*1: Voltage measurement influence includes the offset voltage indicated in the basic specifications.
*2: Combination measurement influence provided for use with the IM3590 are offered as reference values.
Reference values describe a specific example and do not constitute a guarantee.
11
Multiplexer Module
Содержание SW1001
Страница 2: ......
Страница 42: ...38 Precautions for Measurement...
Страница 48: ...44 Resetting Scan Operation...
Страница 50: ...46 Checking the Device Status...
Страница 52: ...48 Initialization Settings...
Страница 118: ...114 Sample Programs 3 Select File Save All 4 Check the save location and click Save as...
Страница 156: ...152 Outline Drawings 12 4 Outline Drawings SW1001 Unit mm Tolerance 0 2...
Страница 157: ...153 Outline Drawings SW1002 Unit mm Tolerance 0 2 12 Appendix...
Страница 158: ...154 Outline Drawings...
Страница 161: ...16 01 EN...
Страница 162: ......
Страница 163: ......
Страница 164: ......