34
Measurement between Two Instruments
3
Repeat channel selection and measurement using BT3562 for 8 channels (internal resistance
measurement).
4
Set the connection method to 2-wire.
The signal is connected to
TERMINAL 1
.
5
Repeat channel selection and measurement using DM7276 in the same way as for internal
resistance measurement (OCV measurement).
The sense wire for 4-wire CH1 to CH8 corresponds to 2-wire CH12 to CH19.
6
Set the connection method of
SLOT 2
to 2-wire.
Use
SLOT 2
for external potential for batteries since the measurement points are different.
The signal is connected to
TERMINAL 1
.
7
Repeat channel selection and measurement using DM7276 (external potential measurement
for batteries).
Control example using communication command
[SW1001] :SYST:MOD:WIRE:MODE 1,WIRE4
Set
SLOT 1
to 4-wire (connect to
TERMINAL 2
).
[SW1001] :CLOSE 101
Select
SLOT 1
, CH1.
[SW1001]
∗
OPC?
Check that the channel relay has been closed.
[SW1001] 1
Receive a response “1” to the
∗
OPC?
query.
[BT3562] :READ?
Execute single measurement using BT3562.
[BT3562] 1.0258E-3
Receive measured values.
[SW1001] :CLOSE 102
Select the next CH2.
...
(Repeat until CH8.)
[SW1001] :SYST:MOD:WIRE:MODE 1,WIRE2
Set
SLOT 1
to 2-wire (connect to
TERMINAL 1
).
[SW1001] :CLOSE 112
Select
SLOT 1
, CH12 (4-wire CH1 sense).
[SW1001]
∗
OPC?
Check that the channel relay has been closed.
[SW1001] 1
Receive a response “1” to the
∗
OPC?
query.
[DM7276] :READ?
Execute single measurement using DM7276.
[DM7276] +03.00
Receive measured values.
[SW1001] :CLOSE 113
Select the next CH13 (4-wire CH2 sense).
...
(Repeat until CH19.)
[SW1001] :SYST:MOD:WIRE:MODE 2,WIRE2
Set
SLOT 2
to 2-wire (connect to
TERMINAL 1
).
[SW1001] :CLOSE 201
Select
SLOT 2
, CH1.
[SW1001]
∗
OPC?
Check that the channel relay has been closed.
[SW1001] 1
Receive a response “1” to the
∗
OPC?
query.
[DM7276] :READ?
Execute single measurement using DM7276.
[DM7276] +00.00
Receive measured values.
[SW1001] :CLOSE 202
Select the next CH2.
...
(Repeat until CH8.)
Содержание SW1001
Страница 2: ......
Страница 42: ...38 Precautions for Measurement...
Страница 48: ...44 Resetting Scan Operation...
Страница 50: ...46 Checking the Device Status...
Страница 52: ...48 Initialization Settings...
Страница 118: ...114 Sample Programs 3 Select File Save All 4 Check the save location and click Save as...
Страница 156: ...152 Outline Drawings 12 4 Outline Drawings SW1001 Unit mm Tolerance 0 2...
Страница 157: ...153 Outline Drawings SW1002 Unit mm Tolerance 0 2 12 Appendix...
Страница 158: ...154 Outline Drawings...
Страница 161: ...16 01 EN...
Страница 162: ......
Страница 163: ......
Страница 164: ......