41
Scan Operation
4.4 Scan Operation
Scanning is started by the SCAN signal of EXT. I/O or the communication
*
TRG
command. The scan
channel is moved forward by the SCAN signal or the
*
TRG
command.
When the channel is the final channel, scanning is completed by the SCAN signal or the
*
TRG
command,
and all relay is opened and goes back to the beginning of the scan list. Note that scan operation will not
complete until another SCAN signal or
*
TRG
command is input in the final channel state.
Scan operation when the trigger source is set to STEP
SCAN signal or
*
TRG
command
Channel
Open
Start
channel
Open
Next
channel
…
Open
Final
channel
Open
Scan complete
CLOSE signal
OFF
ON
The CLOSE signal is output in pulses. The pulse width can be set using the following command
:
IO:PULSe:TIME <0.001
to
0.100/MIN/MAX/DEF>
If the next channel is closed before the CLOSE signal pulse reaches the set pulse width, CLOSE
signal pulse will automatically turn off.
Item
Communication command
Setting
procedure
*
TRG
Starts scanning and proceeds along the scan channels.
:ABORt
Aborts scan operation and returns to the beginning of the scan
channel.
Setting
example
Executes scanning twice from CH1 of
SLOT 1 to
CH22 of
SLOT 3
.
:SCAN 101:322
:TRIG:SOUR STEP
Advances the scan channel one step at a time using the
*
TRG
command (or SCAN signal).
*
TRG
(Repeats for the number of channels.)
*
TRG
Scan operation complete by final channel state.
Scanning is performed again after completion.
*
TRG
(Repeats for the number of channels.)
*
TRG
Scan operation complete by final channel state.
Aborts scan operation and restarts from the beginning.
:SCAN 101:322
:TRIG:SOUR STEP
Advances the scan channel one step at a time using the
*
TRG
command (or SCAN signal).
*
TRG
(Repeats.)
:ABOR
(Aborts scan operation and returns to the beginning of the scan
channel.)
*
TRG
(Repeats for the number of channels from the first channel.)
All channels are open after scanning is stopped.
The following commands are disabled during scanning.
• Setting the connection method
• Setting the shield switching
• Channel switching
• Setting the channel delay
• Setting the scan list
• Setting concerning the EXT. I/O
• Channel control due to CLOSE command
4
Scan Function
Содержание SW1001
Страница 2: ......
Страница 42: ...38 Precautions for Measurement...
Страница 48: ...44 Resetting Scan Operation...
Страница 50: ...46 Checking the Device Status...
Страница 52: ...48 Initialization Settings...
Страница 118: ...114 Sample Programs 3 Select File Save All 4 Check the save location and click Save as...
Страница 156: ...152 Outline Drawings 12 4 Outline Drawings SW1001 Unit mm Tolerance 0 2...
Страница 157: ...153 Outline Drawings SW1002 Unit mm Tolerance 0 2 12 Appendix...
Страница 158: ...154 Outline Drawings...
Страница 161: ...16 01 EN...
Страница 162: ......
Страница 163: ......
Страница 164: ......