13.4 Disposing of the Instrument
200
When disposing of this instrument, remove the lithium battery and
dispose of battery and instrument in accordance with local regula-
tions.
Lithium Battery Removal
Preparation items
13.4 Disposing of the Instrument
• To avoid electric shock, turn off the power switch and dis-
connect the cord and cables before removing the lithium
battery.
• Battery may explode if mistreated. Do not short-circuit,
recharge, disassemble or dispose of in fire.
• Keep batteries away from children to prevent accidental
swallowing.
CALIFORNIA, USA ONLY
This product contains a CR Coin Lithium Battery which contains
Perchlorate Material - special handling may apply.
See www.dtsc.ca.gov/hazardouswaste/perchlorate
Phillips head screwdriver
Tweezers
1
Turn off the power switch.
2
If any cords are connected, for example voltage cords, clamp sensors,
or the AC adapter, disconnect them.
If the PW9002 Battery Set (9459 Battery Pack) is installed, remove it.
See: