10.4 Using the PW3360 Settings and Download Application (USB)
136
The effects of electromagnetic interference such as noise from an
external source may cause communications errors when using a
USB connection. If you encounter such errors, attach a commer-
cially available ferrite clamp to the USB cable as shown in the fig-
ure below. The effectiveness of the clamp can be increased as
follows:
• Place the clamp as close to the computer-side connector as pos-
sible.
• Loop the cable through the clamp several times.
Ferrite core
Position as close to
the computer as
Reference
A ferrite core (segmented core) from Seiwa Electric Mfg. Co., Ltd.,
is shown in the figure.
Model number: E04SR301334