Accuracy Table
5
AUTO V
Range
Accuracy
*1
Input impedance
DC, 40 to 500 Hz
Over 500 Hz to 1 kHz
600.0 V
±2.0% rdg. ±3 dgt. ±4.0% rdg. ±3 dgt.
900 k
Ω
±20%
*2
1800 k
Ω
±20%
*3
• Overload protection: 1100 V DC (1870 V DC
*3
)/1100 V AC or 2×10
7
V • Hz
(energized for 1 minute)
• Transient overvoltage: 8000 V
• Crest factor: The crest factor is 3 up to 4000 counts and reduces linearly to
2 at 6000 counts.
• Connection method: DC coupling
*1: For AC voltage, the accuracy is speci
fi
ed in 1% or more of the range,
however, ±5 dgt. should be added to 5% or less of the range.
• Accuracy guarantee range for frequency: 40 Hz to 1 kHz (Measured
values outside the accuracy guarantee range for frequency are also
displayed.) The accuracy is not speci
fi
ed for strain waveforms outside
the range of 40 Hz to 1 kHz.
• For 100 Hz with the
fi
lter ON, ±1.5% rdg. is added to the accuracy
speci
fi
cation between 40 Hz and 100 Hz and the accuracy is not speci
fi
ed
in 100 Hz or more.
• For 500 Hz with the
fi
lter ON, ±0.5% rdg. is added to the accuracy
speci
fi
cation between 40 Hz and 500 Hz and the accuracy is not speci
fi
ed
in 500 Hz or more.
*2: DT4253, DT4255, DT4256
*3: DT4254