85
Accuracy Table
Minimum sensitivity voltage (sine wave)
Range
Measurement range
AC voltage range
6.000 V 60.00 V
600.0 V
1000 V
99.99
Hz
5.00 Hz to 99.99 Hz
*1
0.600 V
or more
6.00 V
or more
60.0 V or
more
100 V or
more
999.9
Hz
100.0 Hz to 999.9 Hz
0.600 V
or more
6.00 V
or more
60.0 V or
more
100 V or
more
9.999
kHz
1.000 kHz to 9.999 kHz
0.600 V
or more
6.00 V
or more
60.0 V or
more
100 V or
more
99.99
kHz
10.00 kHz to 50.00 kHz
1.800 V
or more
12.00 V
or more
120.0 V
or more
230 V or
more
Over 50.00 kHz to 99.99
kHz
3.000 V
or more
24.00 V
or more
240.0 V
or more
400 V or
more
• The voltage input is up to 2×10
7
V • Hz.
• [----] appears when no measurement can be made.
*1: The measurement range from 5.00 Hz is only for the 6.000 V range. The
measurement range for other voltage ranges is 40.00 Hz to 99.99 Hz.
Minimum sensitivity current (sine wave)
Range
Measurement range
AC current range
600.0 mA
6.000 A
10.00 A
99.99 Hz
40.00 Hz to 99.99 Hz
60.0 mA
or more
0.600 A or
more
3.00 A or
more
999.9 Hz
100.0 Hz to 999.9 Hz
60.0 mA
or more
0.600 A or
more
3.00 A or
more
9.999 kHz 1.000 kHz to 9.999 kHz
60.0 mA
or more
0.600 A or
more
3.00 A or
more
www.