63
Retaining the Measured Value
Conditions for auto holding
Function
Stable range for auto
holding
(Display count)
Threshold for auto
holding
(Dead zone display count)
AC voltage
120 or less (except 1000 V
range)
20 or less (1000 V range)
120 or less (except 1000 V
range)
20 or less (1000 V range)
DC voltage*
2
120 or less (except 1000 V
range*
1
)
20 or less (1000 V range*
1
)
120 or less (except 1000 V
range*
1
)
20 or less (1000 V range*
1
)
AUTO V
120 or less
120 or less
Continuity check 100 or less
4900 or more
Resistance
100 or less
4900 or more
Diode
40 or less
1460 or more
AC (Clamp)
50/100/25/50/100/25/50 or
less (each range)
50/100/25/50/100/25/50 or
less (each range)
DC (
μ
A)
120 or less
120 or less
DC (mA)
120 or less
120 or less
DC (A)
120 or less (except 10 A
range)
20 or less (10 A range)
120 or less (except 10 A
range)
20 or less (10 A range)
AC (A)
120 or less (except 10 A
range)
20 or less (10 A range)
120 or less (except 10 A
range)
20 or less (10 A range)
*1: DT4254 is 1500 V range.