57
Troubleshooting
Problem
Check
Solution
See
Display will not
reset to zero
•
Has DC or AC+DC
measurement mode been
selected?
•
Execute zero adjustment.
•
Is the possible range for
zero adjustment exceeded?
×1 range
Approx.
± 80 dgt.
×10 range
Approx.
± 800 dgt.
•
If the range is exceeded,
the current sensor must be
repaired.
−
•
Is the response speed set to
SLOW
?
•
If the response speed is set
to
SLOW
, the display may
not read zero immediately.
Display value is
lower than expected
Current measurement
•
Are the sensor jaw tips
damaged?
•
Is damage is evident, have it
repaired.
−
•
Are the sensor jaws tightly
closed?
•
Close them tightly. If they
are not properly closed, the
value will be low.
−
•
Is the measurement mode
suitable?
•
Select a measurement
mode that is suitable for the
conductor to be measured.
•
The method for executing
zero adjustment in the DC or
AC+DC measurement mode
is correct?
•
Execute zero adjustment.
A lower than expected
display value will result if
the measurement are made
while the unit is displaying a
negative value.
•
Are you measuring a current
of 10 Hz or lower?
•
To measure current of not
more than 10 Hz, select the
AC+DC mode, and set the
display/output update rate to
SLOW
.
•
Has
SLOW
been set with
auto ranging?
•
With
SLOW
, measurement
of rapid variation is not
possible. Set
NORMAL
or
FAST
.
10
9
8
7
6
5
4
3
2
1
Appx.
Ind.
Содержание CM7290
Страница 1: ...CM7290 CM7291 Instruction Manual DISPLAY UNIT EN Oct 2018 Revised edition 2 CM7290A961 02 18 10H...
Страница 2: ......
Страница 36: ...32 Connecting to an External Device...
Страница 58: ...54 External Interface Specifications only for CM7291...
Страница 77: ......
Страница 78: ......
Страница 79: ......
Страница 80: ......