58
Troubleshooting
Problem
Check
Solution
See
Display value is
lower than expected
• Is the filter function being
used
• Disable the filter function.
• If components of 60 Hz or
higher are present, the value
will be low.
•
Does the
indicator
flash?
•
The batteries have run low.
Replace with new batteries.
•
Does the frequency of the
current being measured fall
outside the range defined
in the product specifications
(for either the unit or the
current sensor)?
•
Use a current sensor
that conforms with the
specifications.
•
If the inverter's carrier
frequency is high, the display
value will be lower than the
total rms value.
•
Has the peak value
exceeded the product
specification?
•
If possible, use a larger
current range.
•
If that is not possible, use a
current sensor that conforms
with the specifications.
Frequency measurement
•
Are you measuring a special
waveform - of an inverter,
etc.?
•
In frequency measurement,
a frequency value is
calculated from a count of
the number of times that a
particular threshold value is
exceeded. This means that
even in the case of a special
waveform, it may be possible
to measure it by changing
the mode – AC or AC+DC,
etc. – or changing the range.
−
• Is the current input value 300
dgt. or more of the range?
•
Use an appropriate range
and/or an appropriate
current sensor.
Содержание CM7290
Страница 1: ...CM7290 CM7291 Instruction Manual DISPLAY UNIT EN Oct 2018 Revised edition 2 CM7290A961 02 18 10H...
Страница 2: ......
Страница 36: ...32 Connecting to an External Device...
Страница 58: ...54 External Interface Specifications only for CM7291...
Страница 77: ......
Страница 78: ......
Страница 79: ......
Страница 80: ......