21
Tables of Key Operations
Desired operation
(function)
: Enabled (factory
settings)
How to set/
cancel it
Screen display
Enabled: Lit
Disabled: Unlit
Description
Have backlight
turned off
automatically
Turn on
power while
pressing key
−
•
Backlight will be turned off
automatically after being lit
for approx. 40 seconds.
•
Setting is memorized when
power is turned off.
Canceling auto
power save (APS)
operation while
using the batteries
Turn on
power while
pressing key
APS
•
When battery-powered,
power is turned off
automatically if the unit is
not operated for approx.
10 minutes.
•
Before the power is turned
off,
APS
on the screen will
flash and the buzzer will
sound.
•
The APS function will be
automatically disabled if
the output, analysis display
or keylock function is
enabled. (When that function
is disabled, APS will be
automatically re-enabled.)
•
APS is disabled when an
AC adapter is used.
Remove noise and
other unwanted
frequency
components
(Filter function
*1
)
1-second
long press
This can only be set in the
AC, AC+DC, and Hz modes.
*1: The display value and output value are values that have been passed through a low-pass
filter (LPF) with cut-off frequencies (fc) around 180 Hz. With the filter function enabled, the
180 Hz components can be damped by -3 dB (approx. 30%). Use this function to remove
carrier components from the inverter output side (secondary side) or if noise is annoying you,
and in similar cases.
10
9
8
7
6
5
4
3
2
1
Appx.
Ind.
Содержание CM7290
Страница 1: ...CM7290 CM7291 Instruction Manual DISPLAY UNIT EN Oct 2018 Revised edition 2 CM7290A961 02 18 10H...
Страница 2: ......
Страница 36: ...32 Connecting to an External Device...
Страница 58: ...54 External Interface Specifications only for CM7291...
Страница 77: ......
Страница 78: ......
Страница 79: ......
Страница 80: ......