AC INRUSH (Inrush current)
1. AC INRUSH measured value
Range
Accuracy guarantee range
(Resolution)
Accuracy guarantee
frequency range
Measurement accuracy
600.0 mA
60.0 mA to 600.0 mA (0.1 mA)
40 Hz ≤ f ≤ 1 kHz
±5.0% rdg ±1.0 mA
6.000 A
±0.600 A to ±6.000 A (0.001 A)
40 Hz ≤ f ≤ 1 kHz
±5.0% rdg ±0.010 A
60.00 A
±1.00 A to ±60.00 A (0.01 A)
40 Hz ≤ f ≤ 1 kHz
±5.0% rdg ±0.10 A
200.0 A
±10.0 A to ±200.0 A (0.1 A)
40 Hz ≤ f ≤ 1 kHz
±5.0% rdg ±1.0 A
2. AC INRUSH PEAK value
Range
Accuracy guarantee range
(Resolution)
Accuracy guarantee
frequency range
Measurement accuracy
600.0 mA
60 mA to 1800 mA (1 mA)
40 Hz ≤ f ≤ 1 kHz
±6.0% rdg ±10 mA
6.000 A
±0.60 A to ±18.00 A (0.01 A)
40 Hz ≤ f ≤ 1 kHz
±6.0% rdg ±0.10 A
60.00 A
±1.0 A to ±180.0 A (0.1 A)
40 Hz ≤ f ≤ 1 kHz
±6.0% rdg ±1.5 A
200.0 A
±10 A to ±300 A (1 A)
40 Hz ≤ f ≤ 1 kHz
±6.0% rdg ±15 A
(3)
Frequency measurement
Range
(Auto-range threshold value)
Accuracy guarantee range
(Resolution)
Measurement accuracy
999.9 Hz (over 9999 counts)
15.0 Hz to 999.9 Hz (0.1 Hz)
±0.1% rdg ±0.1 Hz
2000 Hz (less than 900 counts)
900 Hz to 2000 Hz (1 Hz)
±0.1% rdg ±1 Hz
HIOKI CM4002A961-02
Содержание CM4002
Страница 2: ...HIOKI CM4002A961 02...
Страница 11: ...3 Introduction HIOKI CM4002A961 02...
Страница 26: ...18 Part Names Part Names Front Rear The illustration shows the CM4003 6 4 1 7 2 9 3 8 5 6 HIOKI CM4002A961 02...
Страница 32: ...24 Display Panel HIOKI CM4002A961 02...
Страница 40: ...32 Inspection Prior to Measurement HIOKI CM4002A961 02...
Страница 100: ...Index 92 HIOKI CM4002A961 02...
Страница 101: ...HIOKI CM4002A961 02...
Страница 102: ...HIOKI CM4002A961 02...
Страница 103: ...HIOKI CM4002A961 02...
Страница 104: ...www hioki com HIOKI CM4002A961 02...