50
Simple event recording function
5
5
Recording in progress
An event with a duration time of less than 400 ms may not be accurately
measured, failing to detect the event.
Maximum value from the recording start
The blinking red backlight will warn you that the maximum value
exceeds the set threshold value.
Presently
measured value
Press. (Pressing the key again
can disable the function.)
To freeze the value
The instrument will reset the maximum value and return to step
4
(
Start measurement
).
* If you fail to press
within 3 s, the instrument will return to step
5
(
During recording
).
To switch the range or change the threshold value
Press
or
during step
4
(
Start measurement
).
Return to step
2
(
Set the range,
Set the threshold value
).
Press
,
, or
, and then,
within 3 s*
Recording stops.
6
6
Stop recording.
To disable the simple event recording function
Cycle the instrument.