70
4.5 Current Integration
Integral Range Setting
10.
Press the
button to select the range for the integral.
Maximum display is 9999 counts.
• If no integral value output is needed, select AUTO ranging.
• The integral range consists of five components for each cur-
rent range (four are common, so there are a total of six range
components).
When the timer is set, a large integrating range is unnecessary
because the time and current range settings are not displayed
even if the
button is pressed.
• If no output settings are made, the integral range can be
changed during integral measurement.
Integral Range Table
Current Rage
Integral Range
Model
9691
20 A
10.000 Ah 100.00 Ah 1000.0 Ah 10.000 kAh 100.00 kAh
--
--
200 A
--
100.00 Ah 1000.0 Ah 10.000 kAh 100.00 kAh 1000.0 kAh
--
Model
9692
20 A
10.000 Ah 100.00 Ah 1000.0 Ah 10.000 kAh 100.00 kAh
--
--
200 A
--
100.00 Ah 1000.0 Ah 10.000 kAh 100.00 kAh 1000.0 kAh
--
Model
9693
200 A
--
100.00 Ah 1000.0 Ah 10.000 kAh 100.00 kAh 1000.0 kAh
--
2000 A
--
--
1000.0 Ah 10.000 kAh 100.00 kAh 1000.0 kAh 10000 kAh