126
6.4 Combined Accuracy
Peak Display Accuracy
6.4.4 Method of Calculation
This section describes the method of calculation by assuming that
measurement is performed in AC A mode using the 9692.
Total error when displayed value is 160.0 A (60 Hz)
Total error = ±
(1.3%rdg. + 0.3 A)
*1
= ±
(2.08 A + 0.3 A)
*2
= ±
2.3 A
True value when 160.0 A is shown: 157.5 A to 162.3 A
*1 See the accuracy at 10 to 66 Hz in AC mode for H range in the
9692+3290-10 combined display accuracy table on page 124.
*2 rdg. error = 160 A x 1.3%
Total error when REC output is 1.2 V (60 Hz) (for L range)
Total error = ±
(1.8%rdg. +10 mV)
*3
= ±
(21.6 mV+10 mV)
*4
= ±
31.6 mV
True value of 1.2 V output: 1.1684 V to 1.2316 V
(11.684 A to 12.316 A in terms of current)
*3 See the REC Output Accuracy at 10 to 66 Hz in AC mode for L
range in the 9692+3290-10 combined REC output accuracy
table on page 124.
*4 rdg. error = 1.2 V x 1.8%
±
(%rdg.+A)
DC
1 to 10 Hz
10 to 45 Hz
45 to 66 Hz
66 to 1 kHz
L
DC
2.0%+4 A
--
AC+DC
2.0%+4 A
2.5%+4 A
2.5%+4 A
2.0%+4 A
2.5%+4 A
AC
--
2.5%+4 A
2.0%+4 A
2.5%+4 A
H
DC
to 2300 A
2.0%+7 A
--
2300 A to
6.5%+7 A
--
AC+DC
to 2300 A
2.0%+7 A
2.5%+7 A
2.5%+7 A
2.0%+7 A
2.5%+7 A
2300 A to
6.5%+7 A
--
6.5%+7 A
--
AC
to 2300 A
2.5%+7 A
2.0%+7 A
2.5%+7 A
2300 A to
6.5%+7 A
--
Содержание 3290-10