29
4.2 Insulation of Internal Circuit
In the 2301-20, the input circuit and alarm output
are insulated from the CAN bus as shown in the
block diagram below. (Withstand voltage: 1.4
kVAC, 50/60 Hz, Response current: 5 mA, 1
minute)
4.2 Insulation of Internal Circuit
CA
N bu
s
Power
Co
nne
ctor
Modu
le bas
e
Withstand voltage of
1.4 kVAC
Ground
Insulation Withstand
Voltage of 1.4 kVAC
Insulation Withstand
Voltage of 1.4 kVAC
ALAR
M
Alarm
output
TE
MP
/HU
M
Mea
sur
emen
t
Input
In
su
la
tin
g
cir
cuit
Analog
circuit
Digital
circuit
Module-to-
module
communica-
tions circuit
The COM terminal of the alarm output terminal is
used for both TEMP and HUM.