1.2 Major Features
14
Action at memory full:
Stop recording (Memory full stop)
Recording Mode
Instantaneous
Value
MAX/MIN/AVE
Instantaneous
Value +
MAX/MIN/AVE
Quantity of
storable data
30000
15000
12000
Recording interval
1 sec.
8.5 hours
4 hours
3 hours
2 sec.
17 hours
8.5 hours
6.5 hours
5 sec.
1.5 days
21 hours
17 hours
10 sec.
3.5 days
1.5 days
1.5 days
15 sec.
5 days
2.5 days
2 days
20 sec.
7 days
3.5 days
2.5 days
30 sec.
10 days
5 days
4 days
1 min.
21 days
10 days
8 days
2 min.
42 days
21 days
17 days
5 min.
106 days
53 days
42 days
10 min.
213 days
106 days
85 days
15 min.
319 days
159 days
127 days
20 min.
426 days
213 days
170 days
30 min.
639 days
319 days
255 days
60 min.
1279 days
639 days
511 days
When the alarm log is ON, the higher the number
of alarms generated, the smaller the recording
period. (Approx. 1 of the data per alarm)