![Hioki 2301-20 Скачать руководство пользователя страница 37](http://html1.mh-extra.com/html/hioki/2301-20/2301-20_instruction-manual_2137086037.webp)
33
5.3 General Specifications
Alarm Output
Alarm output × 2 CH
(Open collector: 30 VDC, 20 mA max.,
with allowable loss of 75 m Wmax.)
Output is turned ON when an alarm (Hi or
Lo) occurs.
Whether to hold the alarm output can be
selected. The reset switch or a command
can be used to reset alarms.
Data Deletion
All items of data are deleted by a command.
New data will be added to the previous data
at the start of recording.
Power Outage
Protection
After recovering from a power outage, the
2301-20 automatically returns to the state
held before the outage.
5.3 General Specifications
Clock Accuracy
±100 ppm (Reference value at temperature
from 0 to 50°C (32 to 122°F) without the com-
munications module)
Backup
Recorded data (saved in flash memory)
Data loss for up to 2 minutes before and af-
ter a power outage may occur.
Communication
Interface
CAN bus
Maximum Rated
Voltage to Earth 33 Vrms, 70 VDC
Rated Supply
Voltage
5 V±0.3 VDC
Maximum Rated
Power
1.4 W
Withstand
Voltage
1.4 kVAC Between input and alarm output,
Input/Output and CAN bus (50/60 Hz, Re-
sponse current 5 mA, one minutes
Dimensions
Approx. 22.5W × 96H × 85D mm (0.89"W ×
3.78"H × 3.35"D) (excluding projections)
Mass
Approx. 120 g (4.2 oz.)
Содержание 2301-20
Страница 1: ...INSTRUCTION MANUAL For 2301 20 HUMIDITY MODULE...
Страница 2: ......
Страница 14: ...Notes on Use 10...
Страница 21: ...17 1 4 Dimension Diagrams 1 4 Dimension Diagrams 96 1 22 5 1 85 1 93 8 1 103 9 1 100 5 1 9 6 0 5 Unit mm...
Страница 22: ...1 4 Dimension Diagrams 18...
Страница 24: ...2 1 Setting the Module ID 20...
Страница 34: ...4 2 Insulation of Internal Circuit 30...
Страница 40: ...5 4 9764 HUMIDIDTY SENSOR Specifications 36...
Страница 42: ...6 2 Service 38...
Страница 43: ......
Страница 44: ......