3.2 Connecting Input/Output Cables
24
3.2.2 Connecting Cables to the ALARM
Terminal (Alarm output)
1.
Use a flat blade screwdriver to loosen the screws
on the terminal block.
2.
Insert a cable for alarm output into the terminal
block, then tighten the screws (at a tightening
torque of 0.25 N•m).
3.
Connect the terminal block to the ALARM terminal.
Connect the cable for CH1 output to CH1 and COM;
connect the cable for CH2 output to CH2 and COM.
Recommended Cable
Cable
Terminal block
Terminal block
ALARM
terminal
Single-wire
:0.14 to 1.5 mm
2
Stranded-wire :0.14 to 1.0 mm
2
AWG
:26 to 16
Cable strip length:5 mm (0.2")