![Hillenbrand Hill-Rom P1440 Скачать руководство пользователя страница 16](http://html.mh-extra.com/html/hillenbrand/hill-rom-p1440/hill-rom-p1440_user-manual_2134279016.webp)
10
4. Pull the Trendelenburg or Reverse
Trendelenburg handle, and press
the Hilow Down arrow control to
lower the bed to the desired position.
NOTE:
To determine the specific degree of
Trendelenburg or Reverse Trendelenburg, refer to the
Trendelenburg gauges located on each side of the bed.
To change the bed position from Trendelenburg or Reverse
Trendelenburg, press the Hilow Up arrow control to raise the bed
to its highest position.
Locking Out the Patient Controls
WARNING:
Electrical component failure may cause the bed to move without any controls
being operated. Locking the patient controls can significantly reduce potential for
unintentional movement. If a patient’s condition is such that injury could result
from unintentional movement, lock out the patient controls. Failure to do so could
result in patient injury or equipment damage.
Use the lockout controls on the foot end
control panel to help keep the patient
from operating the head, knee, or hilow
functions. When necessary, use the
lockout controls for patient safety.
When a control is locked out, its
lockout’s indicator turns on.
handle
Reverse
Trendelenburg
handle
Trendelenburg
gauge
Trendelenburg
Hilow
lockout
Knee
lockout
Head
lockout