1 4 Technical data
55000-00012_002_01 – 1695233 – 25/01/2018
55
Guidelines and manufacturer declaration - immunity to electromagnetic
interference / portable and mobile radio devices
Portable and mobile RF communication devices should be used no closer to any part of the TruLight™ 1000, including cables,
than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
TruLight™ 1000 is not intended for use in the vicinity of HF surgical devices or in RF shielded rooms for MRI. It was
only tested for interference immunity against radiated fields in the electromagnetic environment mentioned
below, which corresponds to a professional or domestic healthcare environment.
NOTE
Immunity test
EN / IEC 60601-1 test level
Compliance level
Conducted RF interference variables
according to EN / IEC 61000-4-6
3 V
0.15 MHz – 80 MHz
6 V
in the ISM band and amateur radio
bands between
0.15 MHz and 80 MHz
a
)
3 V
0.15 MHz – 80 MHz
6 V
in the ISM band and amateur radio
bands between
0.15 MHz and 80 MHz
a
)
Radiated RF disturbance variables
according to EN / IEC 61000-4-3
10 V/m
80 MHz – 2.7 GHz
10 V/m
80 MHz – 2.7 GHz
a)
= The ISM bands (ISM = industrial, scientific and medical) between 0.15 MHz and 80 MHz are 6.765 MHz to 6.795 MHz;
13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz and 40.66 MHz to 40.70 MHz. The amateur radio frequencies
between 0.15 MHz and 80 MHz are 1.8 MHz to 2.0 MHz; 3.5 MHz to 4.0 MHz; 5.3 MHz to 5.4 MHz, 7 MHz to 7.3 MHz, 10.1
MHz to 10.15 MHz, 14 MHz to 14.2 MHz, 18.07 MHz to 18.17 MHz, 21.0 MHz to 21.4 MHz, 24.89 MHz to 24.99 MHz, 28.0
MHz to 29.7 MHz and 50.0 MHz to 54.0 MHz.