2 Safety instructions
55000-00012_002_01 – 1695233 – 25/01/2018
–
To prevent paint or corrosion damage, only use agents that do not contain
chlorides or halides.
• Please refer to
Chapter 9
for the sterilisation of handles.
• It is essential that the hygiene instructions of the operator are
observed.
Adjustments
Adjusting the device
The manufacturer guarantees the safety and proper working of the
device only under the condition that the work of adjustment has been
done by an authorised hospital technician or a person with equivalent
qualification.
Deinstallation for service purposes
The spring arm may bounce up
If the light head is removed without first moving the spring
arm to the highest limit position, the spring arm will shoot up
and can cause severe injuries:
• The light head may therefore only be de-installed by the technical
customer service
Commissioning
Initial commissioning prior to use
The light system must be handed to the user in a tested state after initial
operation before it can be used in routine medical procedures.
• The initial operation includes functional and safety checks on the
entire lighting system.
• Handover must be documented by a handover declaration.