Troubleshooting
Recurrent faults
If a fault recurs or cannot be remedied, put the device out of service and contact the technical customer service.
Causes of faults and fault removal
NOTE
Error
Possible cause
Remedy
Chapter
Suspension / mobility
Light head drops / rises
Spring force in spring arm is
too strong / too weak
Adjusting the spring force
see Chapter 11.1
Light head moves
with difficulty / easily
Brakes set too stiff / too
loose.
Adjust brakes
see Chapter 11.3
Optical system / light technology
Light intensity too low
Light intensity set too low
Increase light intensity
see Chapter 7.5
Inhomogeneous light field
The light head is outside the
working area
Position the light head
see Chapter 7.7
The light does not light up
Main switch in the
examination room is
switched off
Switch on the main switch
see Chapter 7.6
Light head has been turned
off at the control panel
Press the ON / OFF button
see Chapter 7.6
Mains cable of the mobile
stand version not plugged in
Plug mains cable into socket
see Chapter 6.1.6
Mains cable of the wall-
mounted version not plugged
in
Plug mains cable into socket
see Chapter 6.1.4
Defective electronics
Inform Technical Customer
Service
—
Site power supply is
interrupted
Check site fuses and power
supply
—
Sterilisable handle
Handles are damaged or
show cracks
End of service life has been
reached
Replace handles
see Chapter 7.4
Life span of the sterilisable
handles too short
Wrong sterilisation method
Check sterilisation method
see Chapter 9.3.3