55000-00012_002_01 – 1695233 – 25/01/2018
35
7
Operation
7.1
Working rules
Compliance with safety instructions
The safety instructions must be read and adhered to when using the lighting system
to ensure that it is safely handled and to prevent harm to the patient.
7.2
Preparatory measures
Measures that should be considered before using the lighting system.
Checking the lighting system
Visual inspection
For all versions of the lighting system, a function test and a visual inspection should
be carried out before use, or at least once per week.
Contamination and infection hazard for patients
Loose or damaged parts may fall into wounds. To ensure the safety of
patients, check the components of the lighting system for the following
points before each use:
• Loose parts on light head,
• Visible damage, in particular on the cover plates of the light head
and the sterilisable handle.
• Secure mounting of the sterilisable handle.
---------------------------------------------------------------------------------------
Electric shock
There is a risk of electric shock in the event of contact with
damaged electrical components of the wall-mounted /
mobile stand version:
• Do not connect the lighting system to the mains in the event of
damaged plug connectors or a damaged mains cable.
• Take the device out of service and label it as DEFECTIVE.
---------------------------------------------------------------------------------------
LED failure
After failure of the third LED, the light head no longer achieves the
specified light intensity.
• Take the lighting system out of service.
---------------------------------------------------------------------------------------
Decommissioning
The lighting system is no longer safe to use when the damage described
above or other damage occurs:
• Disconnect the lighting system by using the master switch installed in
the building, or remove the safety connector.
• Secure the master switch or the safety connector against accidental
switch-on / plugging in.
• Label the lighting system as DEFECTIVE!
• Inform Technical Customer Service.