6 Commissioning
55000-00012_002_01 – 1695233 – 25/01/2018
34
Figure 14
6.1.6
Power supply of the mobile stand version
The power supply of the mobile stand version is established
by a mains cable ④ with cold-device plug and a safety
connector.
Electric shock
There is a risk of electric shock in the
event of contact with damaged
electrical components:
• Do not connect the lighting system to the
mains if the plug connectors ③ / ⑤ are
damaged or if the mains cable ④ is damaged.
• Label the device as DEFECTIVE and contact the
technical customer service
Connecting the lighting system
1. Check the cold-device plug ③, the mains plug ⑤ and
the mains cable ④ for damage.
2. Lay the mains cable ④ so that there is no danger of
tripping and so that there is no strain on the cable.
3. Check whether the mains voltage corresponds to the
information on the device label. Ask the local energy
supply company or a specialist electrical company in
cases of doubt.
Earthing of the mains socket
In case of an electrical short circuit, earthing
reduces the danger of electric shock:
• The mobile stand version (Protection Class I)
must only be connected to a properly earthed
shockproof plug socket.
4. Plug the cold-device plug ③ into the cold-device socket
② on the right-hand side of the power supply unit ①.
5. Insert the mains plug ⑤ into a correctly installed and
earthed shockproof plug socket.
6. The device is switched to Standby.
WARNING
WARNING