Adjustments
Figure 17
11.1
Specifications
Adjustment work may only be performed by specialised
technical staff of the operating company following
appropriate hospital technician training by Trumpf Medical.
11.2
Adjusting the spring force on the spring
arm
The weight of the light head is compensated by a spring which
is installed in the spring arm ①. The spring force must be
readjusted if the spring arm with the light head sinks down of
its own accord.
The adjustment screw for setting the spring force is installed:
• In the ceiling and wall mounted version: on the joint
between the boom and the spring arm,
• In the mobile stand version: on the joint between the
stand rod and the spring arm.
1. Carefully lever off the cover cap ③ from the joint of the
spring arm using a thin blade.
2. Swivel the spring arm ① to the top end posi on (+35°).
Uncontrolled rotation of the
spring arm
Risk of injury due to uncontrolled
rotation movement!
• The spring arm may jump up if the spring force
is set too high.
• If the spring force is reduced too far, the spring
arm may slip down due to the weight of the
light head.
• Carefully turn the adjustment screw while
continuously monitoring the positional
stability of the rotation movement.
------------------------------------------------------------
Damage to the adjustment mechanism
The adjustment mechanism can be damaged and
the spring arm made unfit for use if the adjustment
screw is screwed in too far.
3. Adjust the spring force by turning the slot-head screw ②
with a screwdriver:
–
Increasing the spring force:
Unscrew the slot-head
screw ② an clockwise.
–
Reducing the spring force:
Screw in the slot-head
screw ② clockwise.
4. Place the cover cap ③ on the joint and press it on.