The Vest
®
Airway Clearance System, Model 205
User Manual (140643 REV 7)
PROPRIETARY AND CONFIDENTIAL DRAFT 19-DEC-11
An Air Hose Comes out of the Air Pulse Generator or
Disposable Garments during Operation
1. Fully disconnect the Air Hoses from the Air Pulse Generator and
the disposable garment.
2. Clean the inside and outside of these (see “Cleaning” on page 54):
•
Ends of the Air Hose
•
Connection Ports on the disposable garment
•
Air Hose Connector Ports outlets on the Air Pulse Generator
3. Connect the Air Hoses to the Air Pulse Generator and disposable
garment (see “Set Up the Air Pulse Generator” on page 12).
4. If the problem is not corrected, call Hill-Rom Technical Support.
For contact information, refer to the back cover.
Screen Shows “RESTARTING”
1. The system has experienced an unexpected event and is restarting.
2. If the problem is not corrected, call Hill-Rom Technical Support.
For contact information, refer to the back cover.