20
The Vest
®
Airway Clearance System, Model 205
User Manual (140643 REV 7)
PROPRIETARY AND CONFIDENTIAL DRAFT 19-DEC-11
2. Look at the settings
on the screen. Ramp
Time Percent is the
portion of the
treatment Time
required to increase
from the initial
settings (
Ramp Start
) to the final settings (
Ramp End
). To adjust
the percent to the prescribed therapy, press the right
Up
or
Down
arrows.
3. Press the
ON
button to advance to the next screen.
4. Review the settings for Frequency and Pressure for the initial
setting, Ramp Start.
a.
To adjust the Frequency setting, press the left
Up
or
Down
arrow buttons until the prescribed treatment shows on the
screen. The Frequency, or Hertz (Hz) may be set between 5 and
20 cycles per second.
b. To adjust the
Pressure setting,
press the middle
Up
or
Down
arrow buttons
until the
prescribed
treatment
Pressure shows on the screen. Pressure may be set between 1
and 10.
5. Press the
ON
button to advance to the next screen.
6. Review the settings on the screen. Frequency and Pressure for the
final setting,
Ramp End
, are shown as treatment Time.
a.
To adjust the Frequency setting, press the left
Up
or
Down
arrow buttons until the prescribed setting shows on the screen.
Frequency or Hertz (Hz) may be set between the Ramp Start
setting and 20 cycles per second.
b. To adjust the Pressure setting, press the middle
Up
or
Down
arrow buttons until the prescribed setting shows on the screen.
Pressure may be set between the Ramp Start setting and 10.