The Vest
®
Airway Clearance System, Model 205
User Manual (140643 REV 7)
PROPRIETARY AND CONFIDENTIAL DRAFT 19-DEC-11
Use the disinfectant as specified in the disinfectant manufacturer's
instructions. Do
not
attempt to disinfect or sterilize the disposable
garment for reuse with more than one patient.
Maintenance
WARNING:
Only authorized persons should service The Vest
®
Airway
Clearance System, Model 205
.
Service by unauthorized persons
could cause personal injury or equipment damage.
Minimal routine maintenance and periodic cleaning are necessary for
The Vest
®
Airway Clearance System, Model 205.
Facilities should do these tests and examinations annually:
• Unplug the Air Pulse Generator from its power source.
• Examine the overall condition of the unit for damage or missing
parts.
• Examine the Power Cord and connector for cuts, scrapes, or other
damage.
• Do electrical safety tests in accordance with facility protocols.
• Clean and disinfect the unit (see “Cleaning” on page 54).
• Connect the Air Pulse Generator to a disposable garment and to an
appropriate power source. Make sure the unit is operational, and that
all functions operate correctly.